Products
Show All >>
Special Gas Chromatography for Spirits Analysis
back
Special Gas Chromatography for Spirits Analysis
Click to enlarge

Special Gas Chromatography for Spirits Analysis

Order Id: CH-GC-06
Categories: Chromatography  Gas Chromatography 

 

Product Description:

GC-2010A gas chromatograph, controlled by computer, is a kind of analysis instrument used in multi-functional laboratory. By applying double-column, two air paths and steady voltage structure, it has separate capillary sampling system with basic configurations of two detectors--FID and TCD. It can be widely used in industries like chemical, pesticides and oil refining as well as departments such as universities, research departments, etc. 

 

Main Features:

1. Double-column, two air path, dual FID detectors, can be installed with two detectors FID, TCD;

2. Can be equipped with one packed column vaporization chamber and a separated capillary system with capillary split / non-split sampler functioned with membrane cleaning and tail blowing system of excellent performance;

3. Large screen LCD, with intuitive and abundant content, convenient to set parameters;

4. It has intelligent door of good performance for cooling with high cooling speed, which is conducive to program temperature lift analysis and near room temperature operation;

5. The column box has the function of three-order temperature-programming;

6. It has the function of power-off protection and can automatically record set parameters;

7. Over temperature protection; when the temperature exceeds the upper limit, it will automatically cut the power and trigger alarm. 

 

Main technical parameters:

1.Temperature control:

Temperature control range: room temperature+ 15 ℃ -399 ℃

Temperature control accuracy: ± 0.1 ℃

2.FID detector:

Sensitivity: M ≤ 2 × 10-11

Linear range: 107

Noise: ≤ 0.02mV

Drifting: ≤ 0.1mV / h

3.TCD detector

Sensitivity: ≥ 2500mV.ml/mg (benzene)

Noise: ≤ 0.02mV